• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. EB EXC GRN 36IN 0 D/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. EB EXC GRN 36IN 0 D/A SH; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number X524H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an incisional hernia surgery procedure on (b)(6) 2022 and suture was used.A needle was inserted to suture the muscle, but it was very hard, and the needle tip was broken.Because the site where the needle tip fell could be confirmed with the eyes, it was immediately taken out.Changed to another product because the needle could not pass through a rigid muscle and completed suturing.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: were x-rays taken to locate the needle piece(s)? the broken tip was able to confirm visually.(the surgeon was chasing the needle tip with his eyes.) the following information was requested, but unavailable: what measures were taken to retrieve the broken piece? no further information is available.Was there any additional tissue damage as a result of searching for the needle piece? no further information is available.What is the lot number? no further information is available.Event related to mw # 2210968-2022-03026.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB EXC GRN 36IN 0 D/A SH
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14219688
Report Number2210968-2022-03025
Device Sequence Number4261548
Product Code GAT
UDI-Device Identifier10705031057470
UDI-Public10705031057470
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX524H
Device Catalogue NumberX524H
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Initial Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-