Model Number 45-20001
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that there was a breakage of the screwdriver tip.All the debris was removed from the patient.
 
Event or Problem Description
It was reported that there was a breakage of the screwdriver tip.All the debris was removed from the patient.
 
Additional Manufacturer Narrative
The reported event was confirmed.The device inspection revealed the following: the visual inspection has shown that the tip of the screwdriver is broken off, no fragment was returned for evaluation.The breakage surface of the device shows relatively smooth surface without any great abrupt features.The shiny areas on the broken region and the absence of plastic deformation indicate that the breakage was instantaneous (brittle fracture) due to excessive torsional or bending load.Furthermore, a hardness test according to rockwell on the returned item as well as a dimensional inspection indicate that the device is in accordance to the values in the material and in the dimensional specifications.Based on the investigation, the root cause was attributed to an user related issue.The failure was probably caused by torsional or bending overload applied to the device during usage.As a reminder, the instructions for use state: "in the course of the operation, repeatedly check that the connections required for precise positioning between the implant and instruments, or between the instruments themselves, are secure.While rare, intra-operative fracture or breakage of instruments can occur.Instruments which have experienced excessive use or excessive force are susceptible to fracture.Instruments should be examined for wear or damage prior to surgery." a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM ASNIS MICRO, CANNULATED SCREWDRIVER, 2.0MM, AO COUPLING
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14220252
Report Number0008031020-2022-00213
Device Sequence Number4212653
Product Code LXH
UDI-Device Identifier04546540534385
UDI-Public04546540534385
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number45-20001
Device Catalogue Number45-20001
Device Lot Number1000340587
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/05/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age17 YR
Patient SexMale
-
-