| Model Number |
24716 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in a non-tortuous and calcified peroneal artery.A 3.0x220x150 sterling balloon catheter was advanced for dilatation.However, during the first inflation at 3 atmospheres for 5 seconds, the balloon ruptured.The device was removed and no device fragments were left inside the patient's body.The procedure was completed with a non-bsc device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole in the balloon 55mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a pinhole in the balloon.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in a non-tortuous and calcified peroneal artery.A 3.0x220x150 sterling balloon catheter was advanced for dilatation.However, during the first inflation at 3 atmospheres for 5 seconds, the balloon ruptured.The device was removed and no device fragments were left inside the patient's body.The procedure was completed with a non-bsc device.No patient complications were reported.
|
| |
|