• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Model Number G01253
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #- k171764.Device evaluated by mfg: device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, following an unspecified procedure, a safe-t-j rosen catheter exchange wire guide frayed.The inner mandril protruded approximately 3-4cm from the tip of the device and cut the finger of the physician handling the device.The physician's finger was reported to be "fine".A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Upon return evaluation of the device, the weld was broken free of the safety wire but was still attached to the coil.
 
Event or Problem Description
Additional information was received 24may2022.The access site was the radial artery.The patient's anatomy was not tortuous, calcified, or scarred.The wire was not altered from its original condition.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Event summary: as reported, following an unspecified procedure, a safe-t-j rosen catheter exchange wire guide frayed.The inner mandril protruded approximately 3-4cm from the tip of the device and cut the finger of the physician handling the device.The physician's finger was reported to be "fine".The access site was the radial artery.The patient's anatomy was not tortuous, calcified, or scarred.The wire was not altered from its original condition.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Investigation ¿ evaluation: a visual inspection of the returned device was conducted.A document-based investigation was also performed including a review of complaint history, device history record (dhr), the instructions for use, and quality control data.The complainant returned one safe-t-j rosen catheter exchange wire guide to cook for investigation.Physical examination of the returned device confirms that the safety is protruding from the coil from a broken weld.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other related complaints from the lot have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precautions: inspect the wire guide for kinks or damage prior to use.Cook concluded that a component failure unrelated to the design or manufacturing of the complaint device caused this incident.Additionally, the information provided upon review of complaint file, device history record, complaint history, and device master record provide objective evidence to support that the device was manufactured to specification.Per the quality engineering risk assessment, no further action is warranted.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key14220466
Report Number1820334-2022-00659
Device Sequence Number4194773
Product Code DQX
UDI-Device Identifier00827002012530
UDI-Public(01)00827002012530(17)270322(10)14614674
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG01253
Device Catalogue NumberTHSCF-35-260-1.5-ROSEN
Device Lot Number14614674
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer05/24/2022
06/24/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date06/17/2022
07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-