| Model Number |
RTU16C |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
Urinary Tract Infection (2120)
|
| Date of Event |
04/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the bever coude hydrophilic intermittent catheters that the patient received were very difficult to get past the patient's prostate and difficult to control.The patient could not grab the tube, so they could not navigate the catheter at all.It was stated that the patient spent the night in the emergency room with a urinary tract infection.It was unknown what medical intervention was provided for urinary tract infection.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.The potential failure mode for this failure could be "catheter too long / outer diameter too large".Therefore, a potential root cause for this failure could be "incorrect extruding parameters".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions are needed.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: to help reduce the potential risk of infection and/or other complications, do not re-use.If discomfort or any sign of trauma occurs, discontinue use immediately and consult your doctor".The ifu also states "do not use the catheter if there is no water inside the inner packaging or if the product is past its expiry date".The ifu also states: "instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra until urine begins to flow.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands".Correction: f h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the bever coude hydrophilic intermittent catheters that the patient received were very difficult to get past the patient's prostate and difficult to control.The patient could not grab the tube, so they could not navigate the catheter at all.It was stated that the patient spent the night in the emergency room with a urinary tract infection.It was unknown what medical intervention was provided for urinary tract infection.
|
| |
|