Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Urinary Tract Infection (2120)
Date of Event 04/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the bever coude hydrophilic intermittent catheters that the patient received were very difficult to get past the patient's prostate and difficult to control.The patient could not grab the tube, so they could not navigate the catheter at all.It was stated that the patient spent the night in the emergency room with a urinary tract infection.It was unknown what medical intervention was provided for urinary tract infection.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.The potential failure mode for this failure could be "catheter too long / outer diameter too large".Therefore, a potential root cause for this failure could be "incorrect extruding parameters".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions are needed.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: to help reduce the potential risk of infection and/or other complications, do not re-use.If discomfort or any sign of trauma occurs, discontinue use immediately and consult your doctor".The ifu also states "do not use the catheter if there is no water inside the inner packaging or if the product is past its expiry date".The ifu also states: "instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra until urine begins to flow.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands".Correction: f h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the bever coude hydrophilic intermittent catheters that the patient received were very difficult to get past the patient's prostate and difficult to control.The patient could not grab the tube, so they could not navigate the catheter at all.It was stated that the patient spent the night in the emergency room with a urinary tract infection.It was unknown what medical intervention was provided for urinary tract infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERMITTENT CATHETER
Common Device Name
INTERMITTENT CATHETER
Manufacturer (Section D)
HANGZHOU BEVER MEDICAL DEVICES CO., LTD.
no. 8-1, longquan road
cangqian town, yuhang district
hangzhou, zhejiang 31112 1
CH  311121
MDR Report Key14220676
Report Number1018233-2022-03068
Device Sequence Number14799787
Product Code GBM
UDI-Device Identifier06948796204001
UDI-Public(01)06948796204001
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K192468
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
Report Date (Section F) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/31/2023
Device Model NumberRTU16C
Device Catalogue NumberRTU16C
Device Lot NumberT11416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/01/2022
Event Location Other
Date Report to Manufacturer04/11/2022
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/30/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-