• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMUR CEMENTED STANDARD; PROSTHESIS KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMUR CEMENTED STANDARD; PROSTHESIS KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Medical product: tibia fixed cemented catalog # 42542006701 lot # 65262805, psn 2.5mm female scr 25mm 2pk catalog # 42-5099-025-25 lot # 65259757, headless trocar drill pin 75mm catalog # lot # catalog # 00-5901-020-00 lot # 65245936, 3.5mm hex headed screw x 48mm catalog # 00-5901-035-48 lot # 65307237, psn all poly pat ply 35mm catalog # 42-5400-000-35 lot # 65271003, pers rev 3mm off stem 15x135 catalog # 42-5603-135-15 lot # 64799486, pers rev 3mm off stem 16x135 catalog # 42-5603-135-16 lot # 64891211, psn rev tib fix np sz d r catalog # 42-5420-067-02 lot # 65114284, psn rev fem cmt ccr std sz 3r catalog # 42-5046-054-02 lot # 64867229, psn asf cps 18mm ve r 3-5 cd catalog # 42-5226-004-18 lot # 64917506.Reporter had indicated that product will not be returned.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during a revision procedure, the surgeon mistakenly opened left implants which were implanted into the patient.The error was noticed after impacting the components.All components were removed without additional bone loss or complications, and the correct right sided implants were placed.Attempts to obtain additional information have been made; however, no more is available.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h1, h2, h3, h6, h10 reported event was confirmed by review of medical records provided.Device was not returned.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported issue is attributed to user error.Medical records review indicates that left femoral and left tibial implants assembled and implanted, noticed it did not look correct, quickly removed the left sided implants without loss of bone and the right sided femoral and tibial components were implanted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED STANDARD
Common Device Name
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14220842
Report Number0001822565-2022-01241
Device Sequence Number11868729
Product Code JWH
UDI-Device Identifier00889024562516
UDI-Public(01)00889024562516(17)300912(10)64862190
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K191625
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42504605401
Device Lot Number64862190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexFemale
-
-