• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Medical Device Problem Code Failure to Zero (1683)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device returned to manufacturer: the occ cable, tip, device shaft, and sensor port were visually and microscopically examined for damage or any irregularities.Inspection revealed that there were numerous kinks/bends throughout the body of the wire and the tip was stretched.The distal end of the shaft was detached/fractured 29cm proximal of the tip.The detached ends were ovaled, indicating that the device was kinked prior to detachment.Device analysis was conducted by inspecting the sensor by viewing the sensor port to verify that the sensor was in the correct location.There were body fluids in the sensor housing and the sensor appeared to be inside the sensor housing detached from the fiber optic cable.The wire was gently shaken to see if the sensor would move within the sensor housing and the sensor did move which verifies the sensor was detached from the fiber optic.
 
Event or Problem Description
Reportable based on analysis completed on (b)(6) 2022.It was reported that the comet pressure guidewire was not recognized and zeroed.There were no patient complications reported.However, returned device analysis revealed a fractured device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Common Device Name
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14221833
Report Number2134265-2022-04797
Device Sequence Number14822649
Product Code DXO
UDI-Device Identifier08714729904397
UDI-Public08714729904397
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151610
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/27/2023
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0027392894
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/22/2022
Initial Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-