MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC VISUALASE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC VISUALASE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 9735542
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other relevant device(s) are: product id: 9735979; product id: 9736317.No parts were returned for product analysis.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a thermal therapy system being used outside of a procedure.It was reported that the main user password expired, however when the manufacturer representative(rep) went to change the password in maint, they were unable to login with the maint password.The rep was also unable to login to root.The system did not give the rep any "incorrect password" notifications, it looked like it attempted to login then kicked them back to the main screen.There was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: a medtronic representative went to the site to perform a system check out and they replaced the computer.The system functioned as intended.B01, c02, d02 are applicable to the system checkout.The computer was returned for product analysis.The representative was able to log on as root.The system gave a message "password expired, change password".The representative was able to enter a new password and access the system software.B01, c19, d14 are applicable to the product analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUALASE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14222738
Report Number1723170-2022-00638
Device Sequence Number5480066
Product Code GEX
UDI-Device Identifier00763000402082
UDI-Public00763000402082
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K081656
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9735542
Device Catalogue Number9735542
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/22/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1