| Model Number |
115-222 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Seroma (2069)
|
| Date of Event |
03/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No contact information was provided for the initial reporter, therefore additional event, product, and/or patient details are not attainable.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported left side rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of seroma is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side seroma.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified, underweight, broken shell, missing a piece of shell (0-25%), red particles on the shell, wear abrasion, and crease flat.A microscopic analysis was performed which identified, sharp broken on posterior.A dimension measurement in the shell was performed which identify, the thickness within specification.Based on the device analysis, the final assessment is: sharp broken on posterior assessed, as unidentified (tear) opening.Missing shell assessed, as inconclusive.
|
| |
|
Event or Problem Description
|
|
Patient reported, left side rupture.Healthcare professional reported, left side seroma.Device has been explanted.
|
| |
|