MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE HIGH; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3542712 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a (b)(6) caucasian female patient who underwent breast augmentation revision surgery with a 350cc mentor siltex contour profile high breast implant experienced left sided deflation post procedure.As a result, patient had an explantation on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On may 4, 2022 the mentor failure analysis lab has received the device for evaluation.The investigation of the returned device was completed by the failure analysis lab on may 11, 2022.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the sal silt contr diap pkg 350cc breast implant was found to have a tear on the union between shell and patch.A microscopic examination was performed and the cause of the deflation could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.A second product was received.No adverse events were reported for this concomitant (contralateral) device, therefore no further analysis is required.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|