• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. UNKNOWN PROFEMUR STEM; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. UNKNOWN PROFEMUR STEM; HIP COMPONENT Back to Search Results
Model Number PPW001XX
Device Problem Corroded (1131)
Patient Problems Pain (1994); Loss of Range of Motion (2032)
Event Type  Injury  
Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event Description
Allegedly, the patient presented with pain and lack of mobility.The device failed due to corrosion at the neck-stem junction.At revision surgery, it was discovered that the device failed due to corrosion of the oblong taper of the profemur cocr modular neck where it seated in the pocket of the profemur titanium stem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN PROFEMUR STEM
Type of Device
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key14226990
MDR Text Key290292430
Report Number3010536692-2022-00148
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberPPW001XX
Device Catalogue NumberPPW001XX
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/08/2022
Date Manufacturer Received04/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-