Model Number 24681
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the balloon ruptured.A charger balloon was selected for use to treat the iliac artery.The target vessel was 70% stenosed and mildly calcified.During the first inflation at 12 atm, the balloon ruptured.The balloon was exchanged for another balloon.The procedure was completed.There were no patient complications.
 
Additional Manufacturer Narrative
Device eval by mfr: a visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was present in the balloon.A longitudinal tear was identified in the balloon material.The tear measured 15 mm in length and extended from a position 3mm proximal of the distal markerband to 18 mm proximal of the distal markerband.A visual examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.
 
Event or Problem Description
It was reported that the balloon ruptured.A charger balloon was selected for use to treat the iliac artery.The target vessel was 70% stenosed and mildly calcified.During the first inflation at 12 atm, the balloon ruptured.The balloon was exchanged for another balloon.The procedure was completed.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARGER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14228382
Report Number2134265-2022-04543
Device Sequence Number19277089
Product Code FGE
UDI-Device Identifier08714729804765
UDI-Public08714729804765
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K112697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/22/2022
Device Model Number24681
Device Catalogue Number24681
Device Lot Number0024474348
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer05/06/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-