DAVOL INC., SUB. C.R. BARD, INC. -1213643 ENDO-FLO IRRIGATOR; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
0037970 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As reported, the bard/davol endo-flo irrigator leaks fluid when used with a reusable metal non-bard/davol trumpet valve instrument, and there is no leaking when the endo-flo irrigator is being used with disposable plastic instruments.A photo was provided that confirms a fluid leak at the luer lock location where the reusable metal non-bard/davol trumpet valve instrument connects to the irrigation tubing of the bard/davol endo-flo.Neither the bard/davol endo-flo, or the reusable metal non-bard/davol trumpet valve instrument has been provided for evaluation; therefore the potential cause of the incompatibility/leaking between the two devices could not be determined.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Not returned.
|
| |
|
Event or Problem Description
|
|
As reported, during a gallbladder repair procedure on (b)(6) 2022, the bard/davol endo-flo irrigator leaked irrigation fluid when used with a reusable non-bard/davol trumpet valve.There was no reported patient injury.Reporter stated "the tubing does not leak with disposable instruments.It only leaks with the metal instrument.It does not leak when connecting to a plastic connector.When attaching the tubing to the metal irrigation instrument, there is significant leakage at the connector.".
|
| |
|