Model Number MTA4U0
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Iritis (1940)
Date of Event 07/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient experienced anterior chamber acute recurrent iritis, as per surgeon the event caused by mechanical breakdown of iol.The patient had a history of macular degeneration-dry, epi-retinal membrane, non-proliferative diabetic retinopathy and primary open angle glaucoma.The iol was removed in a secondary procedure.The patient issue was resolved, and the patient was not hospitalized.Additional information has been received stated that patient left aphakic and no iol was inserted.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.Both haptics are broken in the gusset area and returned loose in the gauze.Broken haptic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTERIOR CHAMBER PMMA SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14235410
Report Number1119421-2022-00877
Device Sequence Number17245288
Product Code HQL
UDI-Device Identifier00380652535618
UDI-Public00380652535618
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P880087
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model NumberMTA4U0
Device Catalogue NumberMTA4U0.060
Device Lot Number12530523
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer05/05/2022
Initial Report FDA Received Date04/28/2022
Supplement Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age77 YR
Patient SexMale
-
-