• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH BIOLOX® DELTA, CERAMIC FEMORAL HEAD, L, 32/+3.5, TAPER 12/14; BIOLOX DELTA CERAMIC FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH BIOLOX® DELTA, CERAMIC FEMORAL HEAD, L, 32/+3.5, TAPER 12/14; BIOLOX DELTA CERAMIC FEMORAL HEAD Back to Search Results
Model Number N/A
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 03/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that: total hip replacement on right hip done in (b)(6) hospital.Reason for revision is dislocation.Patient involvement-revision.
 
Additional Manufacturer Narrative
Concomitant medical products: medical product: avenir® müller, stem, lateral, uncemented, ha, 4, taper 12/14 catalog#: 01.06010.104; lot#: 2893093.32mm i.D.Size ii elevated rim liner use with 52mm o.D.Size ii shell catalog#: 00-8752-010-32 ; lot#: 63648275.52mm o.D.Size ii porous uncemented with cluster holes shell use with ii liners catalog#: 00-8757-052-01; lot#: 63719117.X-ray received and will be reviewed as part of ongoing investigation.The manufacturer did not receive the device for investigation.The lot number of the device was received.The device history records will be reviewed during investigation.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).Discarded.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.As no product was returned, visual and dimensional evaluations could not be performed.A review of the device history records did not identify any discrepancies that would have contributed to the reported event.Device is used for treatment.There are no recalls or product holds for the reported product.Medical records were not provided and the radiograph does not confirm reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX® DELTA, CERAMIC FEMORAL HEAD, L, 32/+3.5, TAPER 12/14
Common Device Name
BIOLOX DELTA CERAMIC FEMORAL HEAD
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14235452
Report Number0009613350-2022-00259
Device Sequence Number4482028
Product Code LZO
UDI-Device Identifier00889024430334
UDI-Public(01)00889024430334(17)271103(10)2924380
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K071535
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number00-8775-032-03
Device Lot Number2924380
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer07/04/2022
Initial Report FDA Received Date04/28/2022
Supplement Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age66 YR
Patient SexFemale
Patient Weight70 KG
-
-