Model Number N/A
Device Problems Patient Device Interaction Problem (4001); Migration (4003)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2022
Event Type  Injury  
Manufacturer Narrative
Investigation of this incident is currently ongoing.A followup/ final report will be submitted when additional information becomes available.Zimmer reference (b)(4).Concomitant products: blunt tip screw, 4x50mm catalog: 47-2486-050-40 lot: 3024741; blunt tip screw, 4x65mm catalog: 47-2486-065-40 lot: 3039406; cortical bone screw, 4x24mm catalog: 47-2486-124-40 lot: 3078234; cortical bone screw, 4x26mm catalog: 47-2486-126-40 lot: 3078193; proximal humerus nail cap, 10.5x5mm catalog: 47-2488-010-05 lot: 3068499.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2022-00256, 0009613350-2022-00257, 0009613350-2022-00261.
 
Event Description
It was reported that 5 weeks after initial surgery, one of the proximal screws had backed out from the proper position.The surgeon is keeping an eye on the patient condition.No revision surgery will be planned so far.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
Event description: it was reported that operation was performed with ann nail on (b)(6), 2022.After 5 weeks from the initial, surgeon found one of the proximal screws was backed out from the proper position.The surgeon keep an eye on the patient condition as well as no revision will be planned so far.No delay.Patient outcome: migration.Review of received data: - no medical records available.- due diligence: no further due diligence required as all required information to support the conclusion is available or was already requested.Product evaluation: - no product was returned; therefore, visual and dimensional evaluation could not be performed.Review of product documentation: - device purpose: all involved devices are intended for treatment.- product compatibility: the product combination was approved by zimmer biomet.- surgical technique: the surgical technique explains that the locking of the corelock is done using the corelock driver with torque limiting handle.Turn slowly clockwise to tighten and engage the corelock mechanism until a click is felt from the torque limiting handle.- dhr review: the quality records show that all specified characteristics have met the specifications valid at the time of production with no ncr with a potential correlation to the reported event was found.Conclusion: it was reported that operation was performed with ann nail on (b)(6), 2022.After 5 weeks from the initial, surgeon found one of the proximal screws was backed out from the proper position.The surgeon keep an eye on the patient condition as well as no revision will be planned so far.The quality records show that all specified characteristics have met the specifications valid at the time of production.Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition, behaviour or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism as a potential contributing factor.However, as further biomechanical testing was carried out and the performance is in an acceptable range in comparison with legally marketed similar devices, no design changes were conducted.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.The need for corrective measures is not indicated and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event Description
No change to previously reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUNT TIP SCREW, 4X42MM
Type of Device
AFFIXUS NATURAL NAIL SYSTEM HUMERAL NAIL
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14237592
MDR Text Key290289467
Report Number0009613350-2022-00260
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00889024505452
UDI-Public00889024505452
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K181827
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number47-2486-042-40
Device Lot Number3008327
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE NARRATIVE IN H10.
Patient Outcome(s) Other;
-
-