Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that: active articulation bearing was dislocated due to disassociation.The surgeon did not comment on this patient's factors for the dislocation.The modular head and neck were also replaced in this revision surgery.Patient outcome - revision due to dislocation.
 
Additional Manufacturer Narrative
Medical product: modular neck r catalog#: 00-7848-024-00; lot#: 64615781.Ve brg 44mm id28mm catalog#: 110031012; lot#: 65099429.Ml taper kinectiv size9 catalog#: 00-7713-009-00; lot#: 64999726.Bone screw 6.5/15mm catalog#: 00-6250-065-15; lot#: 64492555.Bone screw 6.5/20mm catalog#: 00-6250-065-20; lot#: j6958907.G7 osseoti 54mm multi hole f catalog#: 110010265; lot#: 6987132.G7 dm metal liner 44mm f catalog#: 110024464; lot#: 729500.The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device yet, however it is indicated by complainant that it will be returned for investigation.The lot number of the device was received.The device history records will be reviewed during investigation.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).In transit to manufacturer.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.This device is used for treatment.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h6, h10.Investigation update, the biolox head was returned for investigation.Metal smearing is visible both on the articulation surface and in the inner surface in contact with the stem taper and it is most likely due to revision surgery.Investigation update does not change the outcome of the investigation: a definitive root cause cannot be established.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX® DELTA, CERAMIC FEMORAL HEAD, M, 28/0, TAPER 12/14
Common Device Name
BIOLOX DELTA CERAMIC FEMORAL HEAD
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14239274
Report Number0009613350-2022-00264
Device Sequence Number17242782
Product Code LZO
UDI-Device Identifier00889024430297
UDI-Public(01)00889024430297(17)310730(10)3081667
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K071535
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number00-8775-028-02
Device Lot Number3081667
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer07/08/2022
01/07/2023
Initial Report FDA Received Date04/28/2022
Supplement Report FDA Received Date08/01/2022
01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-