| Model Number |
PT1X |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported a part of their periotome detached in the patient's mouth during a dental procedure.It was reported that the patient was sent to receive x-rays.
|
| |
|
Additional Manufacturer Narrative
|
|
Hu-friedy has requested additional information from the user facility regarding the patient status.However; to date a response has not been received.The periotome subject of the event was returned to hu-friedy for evaluation and the device was confirmed to have fractured near the tip.The evaluation found that the area where the fracture occurred was bent.The observed damage is indicative of excess force being applied to the device over time subsequently resulting in the reported event.The device was manufactured in 2016 making it approximately 6 years old.A follow-up mdr will be submitted should additional information becomes available.
|
| |
|