Model Number 2220J
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Abdominal Pain (1685)
Date of Event 11/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately nine years and two months days post deployment, computed tomography of abdomen and pelvis demonstrated some of the filter struts were extending beyond the confines of the inferior vena cava.Therefore, the investigation is confirmed for the alleged perforation of the inferior vena cava.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 05/2012).
 
Event or Problem Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with pulmonary embolism.Approximately nine years and two months post filter deployment, it was alleged that the filter strut perforated and the patient reportedly experienced abdominal pain.The device has not been removed and there were no reported attempts made to retrieve the filter.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMON NITINOL FILTER SYSTEM
Common Device Name
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14239647
Report Number2020394-2022-00334
Device Sequence Number4338802
Product Code DTK
UDI-Device Identifier00801741040917
UDI-Public(01)00801741040917
Combination Product (Y/N)N
PMA/510(K) Number
K970099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source Other,Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2220J
Device Catalogue Number2220J
Device Lot NumberGFTC2737
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/12/2022
Initial Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/12/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age60 YR
Patient SexFemale
Patient RaceBlack Or African American
-
-