Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately nine years and two months days post deployment, computed tomography of abdomen and pelvis demonstrated some of the filter struts were extending beyond the confines of the inferior vena cava.Therefore, the investigation is confirmed for the alleged perforation of the inferior vena cava.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 05/2012).
|