Model Number A2000
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A facility reported that the "c" clamp on the mayfield 2000 skull clamp (a2000) could not hold the fully fixed position and has a little movement even when the latch is in the closed position.They believe this is a critical fault during surgery.No patient injury or surgical delay has been reported.
 
Additional Manufacturer Narrative
Mayfield 2000 skull clamp (a2000) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.Unit received had the hex bushing replaced due to having too much movement in the lock.Unit was machined to add helicoils to the large starbursts.All worn components were replaced and general maintenance performed.Root cause analysis: complaint confirmed via inspection of the unit.The rocker arm assembly has some movement.The hex bushing was replaced due to the movement of the lock.Probable root cause is routine use and wear.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD 2000 SKULL CLAMP
Common Device Name
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14239720
Report Number3004608878-2022-00075
Device Sequence Number17377969
Product Code HBL
UDI-Device Identifier10381780253655
UDI-Public10381780253655
Combination Product (Y/N)N
PMA/510(K) Number
K932807
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA2000
Device Catalogue NumberA2000
Device Lot Number199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/05/2022
Initial Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-