| Model Number |
A2079 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mayfield infinity xr2 base unit (a2079) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.Unit received had the shock cushion and labels on both handles replaced.The handles were readjusted to the proper pressure.Unit received with the tri-star adapter and 7¿ extension arm.Replacement of worn components and handles readjusted.General maintenance performed.Root cause analysis: complaint confirmed via inspection of the unit.The handles were readjusted to the proper pressure and the shock cushions required replacement.Probable root cause is routine use and wear.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
|
| |
|
Event or Problem Description
|
|
A facility reported that the mayfield infinity xr2 base unit (a2079) had too much movement.The facility has not reported if there was patient involvement, injury or surgical delay related to this event.
|
| |
|