• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL LINEAR STAPLER; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL LINEAR STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Model Number TX60G
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: u5eh06.Device analysis: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tx60g device arrived with the shrouds removed and damaged, the firing trigger broken, and with a reload loaded on the device.The reload was received fully fired and several damaged on the body reload near of retaining pin area.Due to the condition of the device no functional test could be performed.Further analysis shows that the closure trigger was received partially closed and the retaining pin seated on the anvil hole.The trigger was completely closed and it was open when the knob was pressed and it was opened as expected.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: squeeze the closing trigger until a click is heard.The instrument is in an intermediate position, the pin is fully seated in the anvil capturing the tissue, and the jaws are partially open.Reposition tissue within the instrument if desired.Squeeze the closing trigger and the handle fully together until a second click is heard.The closing trigger is now latched to the handle and the jaws are clamped onto the tissue, which is ready to be stapled.The firing trigger will simultaneously move down to the ready-to-fire position.Before firing, check to ensure the retaining pin is seated in the anvil.If the pin is not properly positioned, staples may not form properly, which may result in leakage or disruption of the staple line.Fire the instrument by pulling the firing trigger back completely against the closing trigger until a click is heard, signifying the staples are fully formed.The damage to the firing trigger is possible that the device encounter higher firing forces than normal resulting in a damaged component.As part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure that the device meets the required specifications prior to shipment.A manufacturing record evaluation was performed for the finished device batch/lot number, and no non-conformances were identified.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
 
Event or Problem Description
It was reported that after the release, the device could no longer be opened.Quite a bit of force was used when trying to open it.When the surgeon opened it, the retaining pin from the magazine didn¿t go back.It was then prepared out.No patient harm reported.Procedure: lower lobe resection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
MDR Report Key14240372
Report Number3005075853-2022-02734
Device Sequence Number11846767
Product Code GDW
UDI-Device Identifier10705036003359
UDI-Public10705036003359
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model NumberTX60G
Device Catalogue NumberTX60G
Device Lot NumberU40P4C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/08/2022
Initial Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-