| Model Number |
C408508 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Approximately four to six weeks post procedure, the side port detached from the sheath.The procedure was completed with no issues and the patient returned to their room with the sheath still in place.Approximately, four to six weeks post procedure, the side port of the sheath detached.It was noted that during the time the sheath was left in the patient, the iv would occasionally become stuck on the device.There were no adverse consequences to the patient as a result of the side port detaching and the sheath was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
One image was submitted for evaluation that appeared to show and engage introducer; no device components were returned.Visual inspection revealed the sideport tubing appeared to have been detached from the sheath sideport.Visual inspection was based solely upon a review of the photographs provided.However, information from the field stated that ¿four to six weeks post procedure, the side port of the sheath detached¿.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported sideport detachment could not be conclusively determined.However, the device was left in the patient for an extended period of time.The engage instructions for use (ifu) states, ¿after completing procedure, withdraw the sheath and discard.¿.
|
| |
|