BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Lot Number 0028113884
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2022
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 13-apr-2022.It was reported that crossing difficulties were encountered.The 92% stenosed target lesion was located in the severely tortuous and severely calcified mid left anterior descending artery.A 3.00 x 48mm synergy xd drug eluting stent was advanced for treatment but failed to cross the lesion.The procedure was completed with a different device.No patient complications were reported.However, returned device analysis revealed stent damage.
 
Manufacturer Narrative
Device evaluated by mfr.: a synergy xd mr ous 3.00 x 48mm stent delivery system was returned for analysis.A visual examination of the stent found mid-stent damage, with stent struts listed.The undamaged crimped stent outer diameter was measured using snap gauge and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple hypotube kinks at several locations along the hypotube shaft.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14242230
MDR Text Key290372936
Report Number2134265-2022-04572
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/22/2023
Device Lot Number0028113884
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/13/2022
Initial Date FDA Received04/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexMale
-
-