| Model Number |
405673 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is unknown; awareness date has been used for this field.Address information was not able to be obtained, therefore, nj was used as a place holder.Medical device lot #: 405673 was reported, however, this is not a lot # manufactured for the reported catalog #.Medical device expiration date: unknown.Device manufacture date: unknown.Medical device lot #: 072353a was reported, however, this is not a lot # manufactured for the reported catalog #.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd spinal anesthesia tray 25 gauge 3.5 inch l the medication was ineffective.This occurred 6 times.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the consumer several failed spinals. "we have had several (approx.5) failed spinals using the bupivicaine out of the bd spinal anesthesia trays, "we have another possible failed spinal from a spinal anesthesia tray used today.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary no photos or physical samples that display the reported condition were available for investigation.A device history review could not be completed as no batch number was provided.Based on the available information we are not able to identify a root cause at this time.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd spinal anesthesia tray 25 gauge 3.5 inch l the medication was ineffective.This occurred 6 times.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the consumer several failed spinals. "we have had several (approx.5) failed spinals using the bupivicaine out of the bd spinal anesthesia trays, "we have another possible failed spinal from a spinal anesthesia tray used today.
|
| |
|