Reporter is a j&j employee.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from depuy synthes reports an event in (b)(6) as follows: it was reported that during on-site testing at the in-house distribution center, it was observed that the torque driver device was out of tolerance.According to the report, the device failure measurement was 3.53 nm but the device specification measurement was 2.75 ¿ 3.25 nm.There was no procedure nor patient involvement reported.No additional information was provided.This complaint involves one (1) device.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: a review of the receiving inspection (ri) for x15 torque driver was conducted identifying that lot number gb113525 was released in one batch batch1: released on august 15, 2018 with no discrepancies.Supplier: (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there were no issues on the surface of the x15 torque driver.A dimensional inspection was not performed due to not finding any issue after visual inspection.A functional test was performed.Current and manufactured revisions were used as source of specification.The device is testing high.The complaint condition was replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the x15 torque driver would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed current and manufactured revisions were reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|