| Model Number |
68LP-150 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.The device has been replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis: visual analysis of the returned device identified: crease fold, wear abrasion, and brown particles in fill channel.Leak test and microscopic analysis was performed which identified: a crease sharp opening on radius and observed void >1 mm in non-textured, valve-to shell-bond area.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a crease sharp opening on radius assessed as fold flaw opening.
|
| |
|