MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The patient presented with intracavity stenosis and underwent endovascular repair of arteriovenous fistula.The target lesion was located in the kidney.A 5.0 x40, 75cm gladiator elite balloon catheter was advanced for dilation.However, during inflation, the balloon ruptured after reaching the burst pressure du to highly calcified lesion.The procedure was completed with another of same device.No patient complications were reported and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The patient presented with intracavity stenosis and underwent endovascular repair of arteriovenous fistula.The target lesion was located in the kidney.A 5.0 x40, 75cm gladiator elite balloon catheter was advanced for dilation.However, during inflation, the balloon ruptured after reaching the burst pressure du to highly calcified lesion.The procedure was completed with another of same device.No patient complications were reported and the patient status was stable.It was further reported that the balloon ruptured on the first inflation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14244715
Report Number2134265-2022-04907
Device Sequence Number5015544
Product Code LIT
UDI-Device Identifier08714729809678
UDI-Public08714729809678
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2024
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0026617442
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date04/28/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight71 KG
-
-