Model Number 1002717
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2022
Event Type  malfunction  
Event Description
It was reported that the pump was unresponsive and would not turn back on.The pump was plugged into a power source and the pump successfully turned back on.The pump history was reviewed, and a ¿battery charge start¿ entry was observed prior to the pump shutting down indicating the pump was charging at the time of the shutdown.Customer's blood glucose level was not impacted.
 
Manufacturer Narrative
No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of Device
AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key14244823
MDR Text Key290456637
Report Number3013756811-2022-40548
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K201214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1002717
Device Catalogue Number1002684
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-