• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTALINK BURR; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTALINK BURR; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 3320
Medical Device Problem Codes Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the catheter was fractured.The 85% stenosed target lesion was located in the mildly tortuous and severely calcified left main artery.A 2.00mm rotalink burr and a 330cm rotawire guidewire was selected for use.During the procedure, it was noted that the guidewire was fractured when used along with 2.00mm rotalink burr.According to the physician, the fracture was due to the lesion characteristics.The device was completely removed via normal method and the procedure was completed with another of same device.No complications were reported and patient was stable post procedure.
 
Event or Problem Description
It was reported that the catheter was fractured.The 85% stenosed target lesion was located in the mildly tortuous and severely calcified left main artery.A 2.00mm rotalink burr and a 330cm rotawire guidewire was selected for use.During the procedure, it was noted that the guidewire was fractured when used along with 2.00mm rotalink burr.According to the physician, the fracture was due to the lesion characteristics.The device was completely removed via normal method and the procedure was completed with another of same device.No complications were reported and patient was stable post procedure.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: returned product consisted of the rotablator rotalink burr catheter.The handshake connection, sheath, coil, burr and annulus were visually examined.Inspection of the device found that the coil was kinked and stretched at the handshake connection, but no device fractures were identified.Product analysis did not confirm the reported event, as there were no fracture damages identified to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK BURR
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55311
6515810888
MDR Report Key14245112
Report Number2134265-2022-04848
Device Sequence Number4421251
Product Code MCX
UDI-Device Identifier08714729126881
UDI-Public08714729126881
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/17/2023
Device Model Number3320
Device Catalogue Number3320
Device Lot Number0027155565
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer08/22/2022
Initial Report FDA Received Date04/28/2022
Supplement Report FDA Received Date08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
Patient Weight72 KG
-
-