• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number FT22AN35NGF963N
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a pre-use check the cuff would not inflate.No patient involvement was reported.
 
Additional Manufacturer Narrative
Other, other text: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found the the blockage was due to the teflon tube that bridges the airway line to the lumen in the wall of the shaft created an obstruction.Functional testing found the returned device cuff would not inflate.It is unclear how or when the blockage came into being, however it is highly unlikely that this condition was present prior to leaving the facility due to the two separate leak tests that were confirmed to be passing.The root cause of the issue could not be determined.No corrective actions are planned at this time.As a correction, manufacturing personnel will be alerted to this issue for awareness.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14245994
Report Number3012307300-2022-07123
Device Sequence Number17327572
Product Code JOH
UDI-Device Identifier15019517291773
UDI-Public15019517291773
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFT22AN35NGF963N
Device Catalogue NumberFT22AN35NGF963N
Device Lot NumberST011145
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/22/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer06/13/2022
Initial Report FDA Received Date04/28/2022
Supplement Report FDA Received Date07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/10/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-