| Catalog Number |
07.01728.001 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2022-00120 through 3012447612-2022-00121.
|
| |
|
Event or Problem Description
|
|
It was reported that a virage closure top migrated out of the screw, allowing the rod to move post-operatively.This was discovered via x-ray at the three week follow-up and a revision surgery was performed the next day to remove and replace the screw, rod, and closure top.This is report three of three for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that a virage closure top migrated out of the screw, allowing the rod to move post-operatively.This was discovered via x-ray at the three week follow-up and a revision surgery was performed the next day to remove and replace the screw, rod, and closure top.This is report three of three for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d4: udi number, d9, and h3.Additional information in h4 and h6: component, investigation type, findings, and conclusions.Device evaluation: the returned devices match the information in the complaint file and were examined.There is no damage to the screw or closure top.Root cause: the root cause was unable to be determined.This event could possibly be attributed to the force applied to the construct by patient activities.Dhr review: per dhr review, the parts were likely conforming when they left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|