| Model Number |
1012446-12 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion located in the mid left anterior descending coronary artery that was 90% stenosed.The 2.25 x 12mm nc trek failed to cross the lesion.During removal, resistance was noted.There was no reported adverse patient effect and no clinically significant delay in the procedure.A non-abbott balloon was used to complete the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|