| Model Number |
Z9002 |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the intraocular lens (iol) was inserted into the patient's operative eye; however, it was removed and replaced during the same procedure as there was a scratch on the lens near the haptic.Account provided that the iol was cut out during the removal process.Another lens of same model and diopter was used as the replacement iol.There was no patient injury and no other intervention was required.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts were made to obtain the missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation? yes section d9: returned to manufacturer: yes section d9: returned to manufacturer on: apr 27, 2022 section h3: device evaluated by manufacturer: yes device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the lens was received cut in half.The lens was cleaned and, scratches as well as a tear on the optic body could be observed as well.The reported complaint issue was identified; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|