Model Number 10104-002
Medical Device Problem Code Environmental Particulates (2930)
Health Effect - Clinical Code Eye Burn (2523)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter phone number: (b)(6).Initial reporter fax number: (b)(6).A field service engineer was dispatched to the customer site.The oil mist filter was replaced to resolve the mist/haze issue.Unit meets specifications and was returned to service.Asp complaint ref #: (b)(4).
 
Event or Problem Description
A customer reported an event of a ¿white fog¿ or mist/haze emitting from the sterrad® 100nx sterilizer, and a user experienced an ¿eye burning¿ reaction.An asp field service engineer was dispatched to assess the unit onsite.The eye reaction is considered minor as there was no report of medical treatment; however, this event is being reported as a malfunction report subsequent to a serious injury.
 
Additional Manufacturer Narrative
Additional event information was received that reported a total of three health care workers (hcws) experienced ¿eye irritation¿ for a ¿few minutes¿ and did not require medical treatment.The eye reaction is considered a minor injury as there was no report of medical intervention.H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the mist/haze issue, and system risk analysis (sra).¿ the device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.¿ trending analysis of the mist/haze issue for the sterrad® 100nx unit was reviewed for the prior six months from open date and no significant trend was observed.¿ review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were available for further analysis.The assignable cause of the mist/haze is likely due to the oil mist filter.The asp field service engineer replaced the part and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: cmp-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD® 100NX STERILIZER 2-DR
Common Device Name
STERRAD® EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer (Section G)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
Manufacturer Contact
gabriela mclellan
33 technology drive
irvine, CA 92618
9495030264
MDR Report Key14246644
Report Number2084725-2022-00107
Device Sequence Number14706449
Product Code MLR
UDI-Device Identifier10705037014507
UDI-Public10705037014507
Combination Product (Y/N)N
PMA/510(K) Number
K071385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10104-002
Device Catalogue Number10104-002
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date04/28/2022
Supplement Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/01/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-