Model Number 71175022
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case(b)(4).
 
Event or Problem Description
It was reported that, during an orif of a right tibial plateau fracture, the evos small 2.5mm drill w/ao qc long fractured in the bone.Fragment was unable to be removed.Clinical decision was made that the risks of removing the fragment outweighed the benefits of removing.Surgery was resumed with a back-up device and without any delay.No harm to the patient has been noted beyond what was described.
 
Additional Manufacturer Narrative
H10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the information provided, the clinical root cause of the reported fracture of the evos small 2.5 drill cannot be confirmed.The drill bit is an external communicating device composed of 17-4 ph stainless steel that is neither manufactured nor intended for implantation and should have limited tissue/bone contact as long-term implantation data is not available.Per the complaint description, it was decided that the risks outweighed the benefits of removing the fragment.Since the fragment remains in the patient¿s bone, the potential for micromotion/migration cannot be ruled out.No further clinical/medical assessment is warranted at this time.Should any additional relevant clinical information be provided, this complaint would be re-assessed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL 2.5MM DRILL W/AO QC LONG
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14247408
Report Number1020279-2022-02050
Device Sequence Number11983854
Product Code HWE
UDI-Device Identifier00885556635766
UDI-Public00885556635766
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175022
Device Catalogue Number71175022
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer05/17/2022
Initial Report FDA Received Date04/29/2022
Supplement Report FDA Received Date05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale