Model Number 9550
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2022
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 07apr2022.It was reported that crossing difficulties were encountered.The 3.5mmx20mm, 80% stenosed target lesion was located in the moderately tortuous and severely calcified right coronary artery.A 3.50 x 20 mm promus premier stent was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable post procedure.However, returned device analysis revealed stent damage.
 
Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.A visual examination of the stent found stent damage.Stent struts in the mid-section were lifted from the crimped profile.The undamaged stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks.A visual and tactile examination of the outer and mid shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14247615
MDR Text Key290384080
Report Number2134265-2022-04727
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/29/2023
Device Model Number9550
Device Catalogue Number9550
Device Lot Number0028357957
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight87 KG
-
-