• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN ABRE VENOUS SELF-EXPANDING STENT SYSTEM; STENT, ILIAC VEIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN ABRE VENOUS SELF-EXPANDING STENT SYSTEM; STENT, ILIAC VEIN Back to Search Results
Model Number AB9U16060090
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
Physician used an abre self-expanding stent during treatment of a lesion in the patient's left common iliac vein.There was no damage noted to packaging (i.E.Shelf carton, hoop/tray).No issues were noted when removing the device from the hoop/tray.The device was prepped as per ifu without issue.The device was not passed through a previously deployed stent and no resistance was encountered during advancement.Deployment issues are reported.The thumbscrew/lock-pin was checked for securement prior to procedure and removed prior to deployment of the stent.The stent was attempted to reposition pulling stent back slightly and it deployed on its own.There was no attempts made to remove the stent.There was no deformation noted to the deployed stent.The delivery system was safely removed.There was no vessel damage noted.A second stent was placed to extend from the first one.No patient injury reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABRE VENOUS SELF-EXPANDING STENT SYSTEM
Common Device Name
STENT, ILIAC VEIN
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14247871
Report Number2183870-2022-00162
Device Sequence Number11861757
Product Code QAN
UDI-Device Identifier00643169796331
UDI-Public00643169796331
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P200026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2025
Device Model NumberAB9U16060090
Device Catalogue NumberAB9U16060090
Device Lot NumberB355798
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Initial Report FDA Received Date04/29/2022
Date Device Manufactured03/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-