• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION POWERFLEX EXTREME 6X4 80CM; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION POWERFLEX EXTREME 6X4 80CM; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4156040S
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, during inflation of a 6mm x 4cm 80cm powerflex extreme percutaneous transluminal angioplasty (pta) balloon catheter with a syringe, the balloon was minimally inflated, and a leakage of contrast was observed coming from the balloon.As a result, the 6mm x 4cm 80cm powerflex extreme pta balloon catheter was removed, and an unknown balloon catheter was used to complete the procedure.There was no reported injury to the patient.The 6mm x 4cm 80cm powerflex extreme pta balloon catheter was prepped in the sterile field per the instructions for use (ifu).A 50/50 contrast and saline mixture was used to prep the balloon of the 6mm x 4cm 80cm powerflex extreme and the device maintained negative pressure during preparation.There was no difficulty removing the device from the hoop, removing the protective balloon cover, the stylet, or any of the sterile packaging components and the device was not kinked or damaged in any way prior to being inserted into the patient.There was no resistance or friction experienced as the device was being inserted.The powerflex extreme was advanced over unknown guidewire to treat the lesion and was able to track toward the lesion and cross without issue.The device was never put into an put bend during this procedure and it did not kink during its use.After the balloon leakage occurred, the device was able to be removed easily from the patient and remained in one piece during it removal.Information regarding the patient, the target lesion, target lesion characteristics, contrast used, device used to complete the procedure, and the guidewire used for the procedure was requested but not provided.The device was discarded and will not be returned.
 
Additional Manufacturer Narrative
Complaint conclusion: as reported, during inflation of a 6mm x 4cm 80cm powerflex extreme percutaneous transluminal angioplasty (pta) balloon catheter with a syringe, the balloon was minimally inflated, and a leakage of contrast was observed coming from the balloon.As a result, the 6mm x 4cm 80cm powerflex extreme pta balloon catheter was removed, and an unknown balloon catheter was used to complete the procedure.There was no reported injury to the patient.The 6mm x 4cm 80cm powerflex extreme pta balloon catheter was prepped in the sterile field per the instructions for use (ifu).A 50/50 contrast and saline mixture was used to prep the balloon of the 6mm x 4cm 80cm powerflex extreme and the device maintained negative pressure during preparation.There was no difficulty removing the device from the hoop, removing the protective balloon cover, the stylet, or any of the sterile packaging components and the device was not kinked or damaged in any way prior to being inserted into the patient.There was no resistance or friction experienced as the device was being inserted.The powerflex extreme was advanced over unknown guidewire to treat the lesion and was able to track toward the lesion and cross without issue.The device was never put into an put bend during this procedure and it did not kink during its use.After the balloon leakage occurred, the device was able to be removed easily from the patient and remained in one piece during it removal.Information regarding the patient, the target lesion, target lesion characteristics, contrast used, device used to complete the procedure, and the guidewire used for the procedure was requested but not provided.The product was not returned for analysis as it was discarded.A product history record (phr) review of lot 82220481 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon leakage - during positive pressure¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown and procedural factors likely contributed to the reported event.However, with the limited amount of information available regarding lesion characteristics and without the return of the device for analysis it is difficult to draw a clinical conclusion between the device and the event reported.According to the safety information in the instructions for use ¿balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in-vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEX EXTREME 6X4 80CM
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key14249948
Report Number3009018440-2022-00009
Device Sequence Number14524566
Product Code LIT
UDI-Device Identifier20705032003961
UDI-Public(01)20705032003961(17)240731(10)82220481
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K032737
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number4156040S
Device Catalogue Number4156040S
Device Lot Number82220481
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date04/29/2022
Supplement Report FDA Received Date05/11/2022
Date Device Manufactured04/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK BALLOON; UNK WIRE
Patient SexUnknown
-
-