Model Number 10742326
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A hospital contact name and direct phone number were not provided to siemens.The event was reported by a siemens employee.Siemens completed a technical investigation of the reported event.The root cause is attributed to a malfunctioning x-ray tube.The system has been repaired by the replacement of the malfunctioning x-ray tube.A general product issue was not identified.The material consumption in relation to the installed base is monitored by the capa process.Material consumption values for the last three (3) months are provided in the following for material number 10414460 - vectron x-ray tube: february 2022: 4%, january 2022: 4.18%, december 2021: 4.02%.No further investigation within the complaint process is necessary.No further investigation is deemed necessary.
 
Event or Problem Description
It was reported to siemens that the ct scanner stopped mid-scan during the examination of a (b)(6) female patient.An error was displayed.The customer could not complete the full scan and the scan had to be repeated.Aside from the additional x-ray dose administered to the patient during the successfully completed second scan, no patient injury has been reported to siemens.This report has been filed with an abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM FORCE
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forchheim, 91301
GM   91301
Manufacturer Contact
rebecca tudor
40 liberty blvd.
65-1a
malvern, PA 19355
4843234198
MDR Report Key14250179
Report Number3004977335-2022-23847
Device Sequence Number17242801
Product Code JAK
UDI-Device Identifier04056869006956
UDI-Public04056869006956
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K190578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Repair
Type of Report Initial
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10742326
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2022
Initial Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age4 YR
Patient SexFemale
-
-