ARJOHUNTLEIGH POLSKA SP. Z O.O. ENTERPRISE 9000X; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number 9611AF101BCUSB
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is in progress.The conclusions will be provided within follow-up report once the investigation is completed.
 
Event or Problem Description
Following the information provided the bed was intermittently lowering unintended, when the button was pressed, the bed did not stop moving once the button was released.The arjo engineer attended the bed for evaluation and repaired the bed by replacing the safety side with the built-in controllers.No patient was involved, no injury was sustained.
 
Event or Problem Description
Following the information provided the whole bed platform was intermittently lowering unintended.When the button on the control panel was pressed, the bed did not stop moving once the button was released.No patient was involved, no injury was sustained.The arjo engineer attended the bed for evaluation and repaired the bed by replacing the safety side with the built-in controllers and the bearing plates.The device was confirmed to be working in accordance to specifications after the repair.
 
Additional Manufacturer Narrative
Based on the available information the root cause of the complained scenario cannot be established.As per the information provided, since the bed would not stop lowering after releasing the control button, it is most likely that the issue was related with the failure of the button on control panel, which is built into the safety side of the bed.It is unknown what could have caused the malfunction of the part.The instructions for use for enterprise 9000x bed (746-591-en rev.12) includes the following information related to the subject of this investigation: - "it is recommended to use the function lockout facility on the attendant control panel to prevent unintended bed movement in situations where objects may press against the patient's controls." - "function lockout can be used to prevent operation of the controls, e.G.When inadvertent movement of the mattress platform could injure the patient." arjo device failed to meet its performance specification since it moved unintended.The device was not used for a patient treatment when the malfunction occurred.This complaint is deemed reportable due to unintended movement of the bed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERPRISE 9000X
Common Device Name
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
ARJOHUNTLEIGH POLSKA SP. Z O.O.
ul. ks. piotra wawrzyniaka 2
komorniki PL-62 052
PL  PL-62052
Manufacturer (Section G)
ARJOHUNTLEIGH POLSKA SP. Z O.O.
ul. ks. piotra wawrzyniaka 2
komorniki PL-62 052
PL   PL-62052
Manufacturer Contact
justyna kielbowska
ks. wawrzyniaka 2
komorniki 62-05-2
PL   62-052
883337089
MDR Report Key14250314
Report Number3007420694-2022-00058
Device Sequence Number4887536
Product Code FNL
UDI-Device Identifier05055982755030
UDI-Public(01)05055982755030(11)190509
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number9611AF101BCUSB
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer03/31/2022
Initial Report FDA Received Date04/29/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/09/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-