• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Medical Device Problem Codes Unable to Obtain Readings (1516); Defibrillation/Stimulation Problem (1573)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Complainant alleged that during biomed testing, the device displayed an unknown "defib failure" and "ecg monitoring failure" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Additional Manufacturer Narrative
The device was returned to zoll singapore for evaluation.The customer report was observed and attributed to the monitor board.The monitor board was recommended to be replaced to resolve the problem.Analysis of reports of this type not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key14250546
Report Number1220908-2022-01509
Device Sequence Number5774425
Product Code MKJ
UDI-Device Identifier00847946006365
UDI-Public00847946006365
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer04/08/2022
Initial Report FDA Received Date04/29/2022
Supplement Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-