Catalog Number 955677
Device Problem Inadequate Ultra Filtration (1656)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/31/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with an ak 98 machine, the device triggered an unspecified alarm and after six minutes into treatment, a reverse ultrafiltration occurred.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was evaluated and repaired on site by a local engineer.A short, simulated therapy was successfully performed.A photograph of the sample was provided for evaluation which showed the triggered alarm.The reported condition was verified.The most likely cause of the condition could be related to uf control out of calibration.The o-ring of degassing chamber, o-ring of quick connector and o-ring of water level sensor were replaced.Calibrated uf cell tuning, and variable flow were performed.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 98 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEDOLLA MODENA MONITORS
via modenese, 66
medolla modena 41036
IT   41036
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14250638
MDR Text Key291952503
Report Number9616240-2022-00007
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414125479
UDI-Public(01)07332414125479
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955677
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/04/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-