Model Number FDS45012
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during flow diverter implantation procedure in a patient with posterior communicating artery aneurysm, the proximal end of the subject flow diverter stent did not open completely and it narrowed like a cone.The physician tried to angioplasty the proximal end of the stent open with a balloon but that didn¿t work.Hence, the physician put another stent on top of the subject device to make the proximal end appose the vessel wall.The procedure was completed successfully.No other clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
The device remains implanted in the patient.
 
Event or Problem Description
It was reported that during flow diverter implantation procedure in a patient with posterior communicating artery aneurysm, the proximal end of the subject flow diverter stent did not open completely and it narrowed like a cone.The physician tried to angioplasty the proximal end of the stent open with a balloon but that didn¿t work.Hence, the physician put another stent on top of the subject device to make the proximal end appose the vessel wall.The procedure was completed successfully.No other clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned for the reported event of stent failed/unable to open¿.H3 other text : the device remains implanted in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.5MM X 12MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14250862
Report Number3008881809-2022-00201
Device Sequence Number5043629
Product Code OUT
UDI-Device Identifier07613327386592
UDI-Public07613327386592
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2023
Device Model NumberFDS45012
Device Catalogue NumberFDS45012
Device Lot Number22510831
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date04/29/2022
Supplement Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NEUROFORM ATLAS (STRYKER); TRANSFORM (STRYKER)
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexFemale
Patient RaceWhite
-
-