| Model Number |
37602 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a healthcare provider, via the manufacturer¿s representative (rep), who reported the left extension was planned to be removed and replaced with a different extension to allow for a full body mri an to alleviate the need for a pocket adapter when implanting a rechargeable neurostimulator (replacing a primary cell) on the left side.During the removal of the extension, the surgeon noticed the wires were exposed when pulling the extension out as the outer coating of the extension wire appeared to shred off of the wires leaving them exposed.The extension was removed and replaced and connected to the new rechargeable neurostimulator and existing right sided lead with all impedances normal and the issue being resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: 7482a40, serial# (b)(4), implanted: (b)(6) 2010, explanted: (b)(6) 2022.Product id: 7482a40, serial/lot #: (b)(4), ubd: 05-sep-2012, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received from the manufacturer¿s representative (rep), which was confirmed with the healthcare provider (hcp), reported the cause of the outer coating longer covering the wiring wasn¿t determined.The reason for ins replacement was due to the battery being low and needing replacement due to this.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|