Model Number 130-03-735
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 04/12/2022
Event Type  Injury  
Event Description
Revision surgery, due to post operative infected total knee removal.
 
Manufacturer Narrative
The reason for this revision surgery was reported as an infection. The previous surgery and the surgery detailed in this event occurred 42 days apart. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The devices were disposed of at hospital and not made available to djo surgical for examination. A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements. There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event. The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery. Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to an infection. There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system. Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial). It is also possible that the patient was not compliant with post-surgical instructions. There are multiple factors that may contribute to an infection that are outside of the control of djo surgical. There are no indications of a product or process issue affecting implant safety or effectiveness. Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOVATION KNEE
Type of DeviceDOMED TRI-PEG PATELLA, 35X9MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14250928
MDR Text Key290404724
Report Number1644408-2022-00568
Device Sequence Number1
Product Code OIY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number130-03-735
Device Catalogue Number130-03-735
Device Lot Number156N1761
Was Device Available for Evaluation? No
Date Manufacturer Received04/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/29/2022 Patient Sequence Number: 1
Treatment
241-02-109 LOT 953R1237; 342-12-709 LOT 123T1026; 352-02-109 LOT 311V1051
-
-