The reported observation of ¿eri, ipg approaching end of life¿ was confirmed.The root cause of the reported observation was due to the device having normal battery depletion.The estimated longevity would have been 2.6 years +/-.25-year per ¿ (b)(4), when using the as received programmed settings and assuming 1000-ohm program impedances, for device model 6662.The total stimulation on time was 2.56 years with the device not declaring end of life (eol), suggesting the device had normal battery depletion at the time of the observation.The device was subjected to a functional test on automated test equipment (ate).This tester verifies the electrical performance of the control/communication circuitry, and output signal integrity.All portions of ate testing passed.Hence, this event is no longer reportable.
|