MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE Back to Search Results
Catalog Number 5000-01-01
Device Problems Defective Component (2292); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2022
Event Type  malfunction  
Event Description
It was reported that the electricity was misguided in the arctic sun device. Per follow up information received on 06apr2022,there was no patient involved and the heater was replaced. Per follow up information received on 07apr2022, it was stated that the inner insolation of the heater was no longer sufficient. A fault current was generated, this problem was with the heater isolation triggered the hospital's iso power fuse. For this reason, the defective heater was replaced. The electricity was not misguided, it was a wear and tear and the production date was 2016.
 
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN® 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14251494
MDR Text Key290445458
Report Number1018233-2022-03136
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number5000-01-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 04/29/2022 Patient Sequence Number: 1
-
-