| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of asetf111 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported "i have a home health nurse who comes to my home to access me & she accessed me tuesday & when i hooked up to my pump is started leaking out my bandage & soaked my shirt.".
|
| |
|
Event or Problem Description
|
|
It was reported "i have a home health nurse who comes to my home to access me & she accessed me tuesday & when i hooked up to my pump is started leaking out my bandage & soaked my shirt.".Add info rcvd: it was reported by patient "i have a bard power port & on tuesday my nurse used batch #asetf111 & it was removed about 4 hrs later on (b)(6) 2022.".
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking infusion set was unconfirmed because the problem could not be reproduced.The product returned for evaluation was one 20ga x 1¿ powerloc safety infusion set.Usage residues were observed throughout the sample.The safety mechanism was engaged and a needleless injection cap was attached to the luer adapter.Microscopic inspection of the sample did not reveal any evidence of current or previous leakage.An attempt to infuse water through the sample using a 12ml syringe revealed the sample to be patent to infusion and aspiration with no observed leaks.No leaks were observed during sustained (>15 seconds) pressurization.No deficiencies were discovered during evaluation of the returned sample.Consequently this complaint is unconfirmed at this time.H3 other text : evaluation findings are in section h.11.
|
| |
|