Catalog Number 791430
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
This report summarizes 2 malfunction events where a midwest phoenix handpiece would not hold burs. No injury resulted in any of the events.
 
Manufacturer Narrative
2 of 2 devices were returned for evaluation. Evaluation of 1 device found normal chuck wear and the turbine needed to be replaced. The device was repaired and returned to the customer. Evaluation of 1 device found chuck wear, lack of maintenance, and lack of lubrication. The device was cleaned of debris and the turbine was replaced and the device was properly lubricated. The device was repaired and returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMIDWEST PHOENIX
Type of DeviceHANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key14251668
MDR Text Key290452465
Report Number9614977-2022-00022
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K131319
Number of Events Reported2
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation
Type of Report Initial
Report Date 04/29/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number791430
Was Device Available for Evaluation? No Answer Provided
Was Device Evaluated by Manufacturer? No Answer Provided

-
-